The transmitter AMS 5200 is the control, support and evaluation unit for the AMS Flue gas probes. It features state-of-the-art micro processor technology. An extensive range of accessories and flue gas probes allows the use of the transmitter AMS 5200 for almost every application.
Currently, a TÜV certification according to European and German environmental regulations is prepared for the AMS 5200.
An accessory for the AMS 5200 is a software for remote service and control. Remote service via Ethernet or Internet is most common in conjunction with continous emission monitoring system, especially for the electronic measuring data processing. Remote service increases the availability of a measuring system, and reduces the cost for service and downtime.
The O2 transmitter AMS 5200 can be integrated into a computer network by wireless ethernet using the bi-directional RS 232 interface on the Transmitter. This allows the direct access of the AMS service engineer to the service and programm settings of the O2 transmitter. Since access to the AMS 5200 is only possible via a direct dialogue between the on-site personnell and the AMS service engineer, the safety and integrity of the computer network is guaranteed at all times.
Technical data
Transmitter | AMS 5200 |
---|---|
Measuring principle | ZrO2 sensor with Pt electrodes |
Application | Soldering machines, OEM application, Gases Industries, Chemical Industries |
Probes | AMS 3211-000 / 500 / 600 / 700 / 860 Ex |
Measuring range | 0 … 25 Vol % |
Analogue signal port | (0) 4 … 20 mA or 0 … 10V, galvanically separated |
Reproducability | +/- 2 % of the measuring value |
Resolution | C(O2) – 0,01 % |
T90-Time | ca. 20 Seconds |
Display | 2* 16 digit, illuminated LCD display Line 1: Display of concentration in ppm or Vol % Line 2: Messages, device status, sensor parameters |
Messages | 1 system message (measuring value yes / no) max. 2 messages configurable as oxygen value, calibration message, measuring value in the range, flow rate |
Digital communication | serial interface RS232 |
Ambient operating temp. | + 5 °C up to + 60 °C |
Power supply | 110 … 230 VAC or 24 VDC |
Protection / Housing / | IP65 / wall mounting housing / ca. 300 x 260 x 130 mm (hxbxt) |
Dimensions | Ex-d housing for Zone 1 / ca. 300 x 250 x 200 mm (hxbxt) |
Weight | 3,5 kg … 10 kg depending on housing |
Options | pneumatic unit for supplying the probes with reference gas and / or calibration gas automatic calibration by using a pneumatic unit |
Version: AMS 5200 E V-2013-07 |