The oxygen analyser AMS 5200 is designed for the measurement of O2 concentrations in Vol. %, making use of a ZrO2 sensor. The AMS 5200 is based on the "AMS 5000" multi application electronic analyser project that was developed in 2009. A modern, microprocessor controlled SMD type electronic guarantee short response and T-90 times.
Due to the logic and self-explaining menu structure of its software, the operator can change settings and calibrate the oxygen analyser with only three keys. AMS offers a software package for the AMS 5200 to allow remote control and service of the analyser.
Technical data
Analyser | AMS 5200 |
---|---|
Measuring principle | ZrO2 sensor with Pt electrodes |
Application | Soldering machines OEM application Gases Industries Chemical Industries |
Measuring range | 0 … 25 Vol % |
Analogue signal port | (0) 4 … 20 mA or 0 … 10V, galvanically separated |
Reproducability | +/– 2 % of the measuring value |
Resolution | 0,01% |
T90-Time | ca. 20 Seconds |
Display | 2* 16 digit, illuminated LCD display Line 1: display of concentration in ppm or Vol % Line 2: messages, device status, sensor parameters |
Messages | 1 system message (measuring value yes / no) max. 2 messages configurable as oxygen value, calibration message, measuring value in the range, flow rate |
Gas connection | inlet / outlet 3 / 6 mm ferrule pack |
Gas sampling | min. 30 Nl/h, max. 50 Nl/h |
Sample pressure (inlet) | min. 1,01 bar abs., max. 10 bar abs. |
Digital communication | serial interface RS232 |
Ambient operating temp. | + 5 °C up to + 60 °C |
Power supply | 110 … 230 VAC or 24 VDC |
Protection / Housing / Dimensions | IP20 / electronics unit 28 TE / 3HU IP65 / wall mounting housing / ca. 300 x 260 x 130 mm (hxbxt) IP54 / portable housing 42 TE / 3 HE IP54 / panel mounting 144 x 144 mm |
Weight | 2,5 kg … 10 kg depending on housing and built-in options |
Options | Pneumatic suction Electronic flow analysis Software for remote display and service max. 4 measuring ranges automatically switching with digital range detection Auto-calibration Pressure regulator Flowmeter manual 5 way valve Particle filter 2-7µm |
Version: AMS 5200 E V-2013-07 |