AMS 5200

The oxygen analyser AMS 5200 is designed for the measurement of O2 concentrations in Vol. %, making use of a ZrO2 sensor. The AMS 5200 is based on the "AMS 5000" multi application electronic analyser project that was developed in 2009. A modern, microprocessor controlled SMD type electronic guarantee short response and T-90 times.

Due to the logic and self-explaining menu structure of its software, the operator can change settings and calibrate the oxygen analyser with only three keys. AMS offers a software package for the AMS 5200 to allow remote control and service of the analyser.

Analyser settings and calibration can be executed from any Microsoft Windows operated PC. 

Numerous useful options for the oxygen analyser AMS 5200 allow the operation in almost every application.

Technical data

AnalyserAMS 5200
Measuring principleZrO2 sensor with Pt electrodes
ApplicationSoldering machines
OEM application
Gases Industries
Chemical Industries
Measuring range 0 … 25 Vol %
Analogue signal port(0) 4 … 20 mA or 0 … 10V, galvanically separated
Reproducability+/– 2 % of the measuring value
Resolution0,01%
T90-Timeca. 20 Seconds
Display2* 16 digit, illuminated LCD display
Line 1: display of concentration in ppm or Vol %
Line 2: messages, device status, sensor parameters
Messages1 system message (measuring value yes / no)
max. 2 messages configurable as oxygen value,
calibration message, measuring value in the range, flow rate
Gas connectioninlet / outlet 3 / 6 mm ferrule pack
Gas samplingmin. 30 Nl/h, max. 50 Nl/h
Sample pressure (inlet)min. 1,01 bar abs., max. 10 bar abs.
Digital communicationserial interface RS232
Ambient operating temp.+ 5 °C up to + 60 °C
Power supply110 … 230 VAC or 24 VDC
Protection / Housing / DimensionsIP20 / electronics unit 28 TE / 3HU
IP65 / wall mounting housing / ca. 300 x 260 x 130 mm (hxbxt)
IP54 / portable housing 42 TE / 3 HE 
IP54 / panel mounting 144 x 144 mm
Weight2,5 kg … 10 kg depending on housing and built-in options
OptionsPneumatic suction
Electronic flow analysis
Software for remote display and service
max. 4 measuring ranges automatically switching with 
digital range detection
Auto-calibration
Pressure regulator
Flowmeter
manual 5 way valve
Particle filter 2-7µm
Version: AMS 5200 E V-2013-07