AMS 6100

The trace analyser AMS 6100 operates on base of an electrochemical sensor. The Analyser AMS 6100 is a micro-processor controlled analyser for the highly accurate measurement of lowest gas traces. Changes in concentrations of < 0,1 ppm with a resolution of 0,1 ppm can be measured.

The available housings are designed for use in general applications. 

Available options:

  • Pressure reducer
  • Particle filter 2-7 µm
  • Manual 5 way valve
  • Manual purge
  • Electronical/pneumatic gas supply pump
  • Electronical flowmeter
  • Auto calibration
  • 2 free adjustable measuring values
  • Different housings

Technical data

Analyser AMS 6100
Measuring principle Electrochemical Sensor
Messuring componentsH2, CO, H2S, …
Application Gases Industries, Chemical Industries
Measuring range 2, automatic selection, digital ID (optional 4)
Analogue signal port 0 (4) … 20mA, or 0…10 V, galvanically separated
Reproducability+/- 2 % of the measuring value
Resolution <0,1 ppm, depending on the H2S concentration
T90-Time ca. 30…40 seconds depending on the measuring cell
Dispaly 2* 16 digit, illuminated LCD display
Messages 2 free adjustable isolated changeover relays
Gas connection inlet / outlet 3 / 6 mm ferrule pack
Gas sampling built-in inlet / valve, flowmeter
Sample flow min. 20 Nl/h, max. 40 Nl/h
Sample pressure (inlet) min. 1,01 bar abs., max. 2 bar abs.
Sample pressure (measuring cell) max. 50 mbar pressure
Digital communicationserial interface RS 232
Ambient operating temp.– 5 °C up to + 45 °C
Relative humidity of the gas 0 … 99 % not condensing
Power supply 24 VDC or 230 VAC
Protection / Housing / Dimensions IP 20 / electronics unit, 84 TE / 3HU
IP 65 / wall mounting housing / ca. 300 x 260 x 130 mm
IP 54 / portable housing 42 TE / 3 HU
Ex-classificationIP 55 wall mounting housing with continuous purging also certified for applications in Ex-Zone2
Weight2,5 – 7 kg
Options bypass and purge valve, manual
measuring range max. 4, automatic selection, digital ID
external sensor
pressure reducer max. 10 bar, out 50 mbar
Ex-classification; in IP 55 wall mounting housing with
continuous purging also qualified for applications in Ex-Zone2
Version: AMS 6100 E V-2013-07