Technical data
| Analyser | AMS 3140 |
|---|---|
| Measuring principle | Electrochemical Oxygen Sensor |
| Application | Gases Industries, Chemical Industries |
| Measuring ranges | max. 2, automatic switch-over, digital range recognition |
| Measuring range | 0 … 10, 0 … 10000 ppmv |
| Analogue signal port | (0) 4 … 20mA, galvanically separated with automatic switch-over and digital range recognition |
| Reproducability | +/- 2 % of measuring range |
| Resolution | 0,01 ppm – C(O2) – 1 ppm depending on O2 concentration |
| T90-Time | ca. 40 Seconds |
| Display | 8 digit LCD Display, illuminated, for measuring values 2* 20 digit LCD Display, illuminated, for operator settings |
| Messages | 2 potential free adjustable messages |
| Gas connection | Inlet / Outlet 3 / 6 mm Ferrule pack |
| Gas sampling | integrated Inlet- / Outlet valve, mechanical flow meter |
| Sample flow | min. 20 Nl/h, max. 40 Nl/h |
| Sample pressure (Inlet) | min. 1,01 bar abs., max. 2 bar abs. |
| Sample pressure (measuring cell | max. 50 mbar Over pressure |
| Digital communication | serial Interface RS 232 |
| Ambient operating temp. | - 5 °C bis + 45 °C |
| Relative Humidity | 0 … 99 % not condensing |
| Power supply | 24 VDC |
| Protection / Housing | IP 65 / Wall mounted housing / |
| Dimensions | 400 x 500 x 200 (HxBxT) |
| Ex-classification | Certification: IBExU10ATEX1136 ATEX II 2(1)G Ex de (ia Ga) IIC T6 Gb |
| Weight | 18 kg |
| Options | Manual By-pass and Purge valve Pressure reducer max. 10 bar, out 50 mbar |
| Version: AMS 3140 E V01-2011-07 |