AMS 3186 DP – dewpoint analyser

The trace oxygen analyser AMS 3186 DP operates on base of an electrochemical sensor. The trace oxygen analyser AMS 3186 DP is a micro-processor controlled analyser for the highly accurate measurement of lowest oxygen traces.

Changes in concentrations of < 0,1 ppm of oxygen with a resolution of 0,01 ppm can be measured. The available housings are designed for use in general applications. The lowest oxygen measuring range is 0 … 1 ppm. To determine the dewpoint of the sample gas, the analyser has be equipped with a dewpoint measuring system in addition to the oxygen measurement.

Available options:

  • bypass and purge valve, manual
  • 2 free adjustable messages
  • electronically flow control
  • electrical/pneumatic gas pump
  • manual 5 way valve
  • particle filter 2–7µm
  • auto-calibration, also with remote control
  • automatic purge of the sensor
  • pressure reducer
  • different housings

Technical data

Analyser AMS 3186 DP
Measuring principle Electrochemical oxygen sensor
Dewpoint: ceramic sensor
Application Gases Industries, Chemical Industries
Measuring range Oxygen:
4, automatic selection, digital ID
0 … 10, 0 … 1000 ppmv
Dewpoint
+20 … -100°C, optional -40…-120°C
0...3.000ppm
Analogue signal portOxygen:
0 (4) … 20mA, galvanically separated,
with measuring range changeover and digital identification
Dewpoint:
0 (4) … 20mA, galvanically separated
Reproducability +/– 2 % of the measuring value
Resolution Oxygen:
0,01 ppm – C(O2) – 1 ppm, depending on the O2 concentration
Dewpoint:
0,1°C – C(Dew) – 1°C, depending on concentration
T90-Time approx. 30…40 seconds O2,
approx. 40 seconds Dewpoint
Display 2* 16 digit, illuminated LCD display
Messages Status, calibration and service request, 1 message each for O2 / Dewpoint
Gas connection inlet / outlet 3 / 6 mm ferrule pack
Gas sampling built-in inlet / outlet valve, flowmeter
Sample flow min. 20 Nl/h, max. 40 Nl/h
Sample pressure (inlet) min. 1,01 bar abs., max. 2 bar abs.
Sample pressure (measuring cell)max. 50 mbar pressure
Digital communicationserial interface RS 232
Ambient operating temp.– 5 °C up to + 45 °C
Relative humidity of the gas 0 … 99 % not condensing
Power supply 230 VAC, 24 VAC
Protection / Housing / DimensionsIP 54 / wall mounting housing
IP 20 / 19“, 3 HE, 473 mm deep, electronic housing
IP 20 / ½ 19”, 3HE, 250 mm deep, electronic cassette
IP 20 / 63 TE, 3 HE, 300 mm deep portable housing
Ex-classificationin IP 55 wall mounting housing with inert gas purge also
qualified for applications in Ex-Zone 2
Weight 5 – 7 kg
Options bypass- and purge valve, manual
measuring range max. bypass- and purge valve, manual
measuring range max. 4, automatic selection, digital ID
electronically flow control<cr>electrical/pneumatic gas pump
manual 5 way valve
particle filter 2-7µm
auto-calibration, also with remote control
automatic purge of the sensor
pressure reducer max. 10 bar, out 50 mbar
Version: AMS 3186 Dewpoint E V-2013-10