The oxygen analyser AMS 5225 is the right choice for applications to monitor the atmospheres of forming gas. It uses an electrochemical sensor.
To house the electrochemical sensors in its oxygen analysers, AMS has developed a series of measuring chambers which are gas-tight sealed. This guarantees leakage-free gas pathways, which is particularly important in Vol. % applications.
Technical data
Analyser | AMS 5225 |
---|---|
Measuring principle | Electrochemical oxygen sensor |
Application | Soldering machines Welding technology Gases Industries Chemical Industries |
Measuring range | 0 … 25 Vol % |
Analogue signal port | (0) 4 … 20 mA or 0 … 10V, galvanically separated |
Reproducability | +/– 2 % of the measuring value |
Resolution | 0,01% |
T90-Time | ca. 20 Seconds |
Display | 2* 16 digit, illuminated LCD display Line 1: display of concentration in ppm or Vol % Line 2: messages, device status, sensor parameters |
Messages | 1 system message (measuring value yes / no) max. 2 messages configurable as oxygen value, calibration message, measuring value in the range, flow rate |
Gas connection | inlet / outlet 3 / 6 mm ferrule pack |
Gas sampling | min. 30 Nl/h, max. 50 Nl/h |
Sample pressure (inlet) | min. 1,01 bar abs., max. 10 bar abs. |
Digital communication | serial interface RS232 |
Ambient operating temp. | + 5 °C up to + 60 °C |
Power supply | 110 … 230 VAC or 24 VDC |
Protection / Housing / Dimensions | IP20 / electronics unit 28 TE / 3HU IP65 / wall mounting housing / ca. 300 x 260 x 130 mm (hxbxt) IP54 / portable housing 42 TE / 3 HE IP54 / panel mounting 144 x 144 mm |
Weight | 2,5 kg … 10 kg depending on housing and built-in options |
Options | pneumatic pump mechanical and electronic flow analysis software for remote display and service Manual 5 way valve max. 4 measuring ranges automatically switching with digital range detection auto-calibration pressure regulator |
Version: AMS 5225 E V-2013-07 |