The basics of the Oxygen Analyser AMS 5100 Z is the development project "AMS 5000". The Trace Oxygen Analyser AMS 5100 Z is designed for the measurement of the lowest Oxygen traces under presence of oxidable components for example in Air separation plants. For this application the Oxygen Analyser AMS 5100 Z is equipped with a catalytic inactive Zircondioxid sensor. Due to the compact design the Oxygen Analyser AMS 5100 Z is available in different housing types.
A long list of options allows to adopt the Oxygen Analyser AMS 5100 Z for almost every application. The software for remote control and remote service with access via the customer computer network further increase the flexibility of the Oxygen Analyser AMS 5100 Z in an application.
Technical data
Analyser | AMS 5100 |
---|---|
Measuring principle | Catalytic inactive ZrO2 probe |
Application | Gases Industries, Chemical Industries |
Measuring range | 0 … 25 Vol % 0 … 10 ppmv |
Analogue signal port | (0) 4…20 mA or 0…10V, galvanically separated |
Reproducability | +/- 2 % of the measuring value |
Resolution | 0,01 ppm – C(O2) – 0,01 % depending on the O2 concentration |
T90-Time | < 20 seconds |
Display | 2* 16 digit, illuminated LCD display 1. Line: display of concentration in ppm or Vol % 2. Line: Messages, device status, sensor parameters |
Messages | 1 System message (measuring value yes / no) max. 1 message configurable as oxygen value, calibration message, measuring value in the range, flow |
Gas connection | inlet / outlet 3 / 6 mm ferrule pack |
Gas sampling | built-in inlet / outlet valve |
Sample flow | min. 20 Nl/h, max. 40 Nl/h |
sample pressure (inlet) | min. 1,1 bar abs., max. 2 bar abs. |
Digital communication | serial interface RS232 |
Ambient operating temp. | + 5 °C up to + 60 °C |
Power supply | 110 … 230 VAC or 24 VDC |
Protection / Housing / Dimensions | IP20 / electronics unit, 1/2 inch, 19 inch / 3HU IP65 / wall mounting housing / ca. 327 x 270 x 326 mm (hxbxt) IP54 / portable housing 42 TE / 3 HU |
Weight | 2,5 kg … 10 kg depending housing and built-in options |
Options | Pneumatic pump Electronic flow control Software for remote control and service Particle filter 2–7 µm Pressure regulator |
Version: AMS 5100 Z E V-2019-06 |
Oxygen measurement
Please note: The cross interference from combustible gases to the Oxygen measurement depends upon the temperature, the gas flow rate and both the concentration of Oxygen and of the combustible component.
Oxygen concentration | 10 ppmV O2 | 100 ppmV O2 |
---|---|---|
Combustibles | Cross interference | Cross interference |
10 ppm H2 | < -0,5 ppm O2 | < -1 ppm O2 |
10 ppm CO | < -0,4 ppm O2 | < -0,6 ppm O2 |
1000 ppm CH2 | not traceable | < -1 ppm O2 |