The basics of the Oxygen Analyser AMS 5100 Z is the development project "AMS 5000". The Trace Oxygen Analyser AMS 5100 Z is designed for the measurement of the lowest Oxygen traces under presence of oxidable components for example in Air separation plants. For this application the Oxygen Analyser AMS 5100 Z is equipped with a catalytic inactive Zircondioxid sensor. Due to the compact design the Oxygen Analyser AMS 5100 Z is available in different housing types.
A long list of options allows to adopt the Oxygen Analyser AMS 5100 Z for almost every application. The software for remote control and remote service with access via the customer computer network further increase the flexibility of the Oxygen Analyser AMS 5100 Z in an application.
Technical data
| Analyser | AMS 5100 |
|---|---|
| Measuring principle | Catalytic inactive ZrO2 probe |
| Application | Gases Industries, Chemical Industries |
| Measuring range | 0 … 25 Vol % 0 … 10 ppmv |
| Analogue signal port | (0) 4…20 mA or 0…10V, galvanically separated |
| Reproducability | +/- 2 % of the measuring value |
| Resolution | 0,01 ppm – C(O2) – 0,01 % depending on the O2 concentration |
| T90-Time | < 20 seconds |
| Display | 2* 16 digit, illuminated LCD display 1. Line: display of concentration in ppm or Vol % 2. Line: Messages, device status, sensor parameters |
| Messages | 1 System message (measuring value yes / no) max. 1 message configurable as oxygen value, calibration message, measuring value in the range, flow |
| Gas connection | inlet / outlet 3 / 6 mm ferrule pack |
| Gas sampling | built-in inlet / outlet valve |
| Sample flow | min. 20 Nl/h, max. 40 Nl/h |
| sample pressure (inlet) | min. 1,1 bar abs., max. 2 bar abs. |
| Digital communication | serial interface RS232 |
| Ambient operating temp. | + 5 °C up to + 60 °C |
| Power supply | 110 … 230 VAC or 24 VDC |
| Protection / Housing / Dimensions | IP20 / electronics unit, 1/2 inch, 19 inch / 3HU IP65 / wall mounting housing / ca. 327 x 270 x 326 mm (hxbxt) IP54 / portable housing 42 TE / 3 HU |
| Weight | 2,5 kg … 10 kg depending housing and built-in options |
| Options | Pneumatic pump Electronic flow control Software for remote control and service Particle filter 2–7 µm Pressure regulator |
| Version: AMS 5100 Z E V-2019-06 |
Oxygen measurement
Please note: The cross interference from combustible gases to the Oxygen measurement depends upon the temperature, the gas flow rate and both the concentration of Oxygen and of the combustible component.
| Oxygen concentration | 10 ppmV O2 | 100 ppmV O2 |
|---|---|---|
| Combustibles | Cross interference | Cross interference |
| 10 ppm H2 | < -0,5 ppm O2 | < -1 ppm O2 |
| 10 ppm CO | < -0,4 ppm O2 | < -0,6 ppm O2 |
| 1000 ppm CH2 | not traceable | < -1 ppm O2 |