AMS 3220

The oxygen analyser AMS 3220 is designed for the measurement of oxygen concentrations in noble gases and process gases without combustible impurities. For this application the oxygen analyser AMS 3220 is equipped with a zircondioxid sensor. Due to the compact design the oxygen analyser AMS 3220 fits in an electronic housing IP20, 28 TE, 3 HE with a depth of 250 mm.

The oxygen analyser AMS 3220 is available in different housings for general applications and classified areas. The oxygen analyser AMS 3220 is certified by the German TÜV according to European and German environmental regulations and fulfils the requirements for QAL 1 according EN 14181 and EN ISO 14956. The oxygen analyser AMS 3220 is tested for suitable flue gas applications according to TALuft, 13. und 17. BImSchV.

Available options:

  • pressure reducer
  • particle filter 2-7µm
  • manual 5 way valve
  • electronical/pneumatic gas supply pump
  • electronical flowmeter
  • auto calibration
  • back purge
  • 2 free adjustable measuring values
  • different housings

Technical data

AnalyserAMS 3220
ApplicationCertified for flue gas applications according
European and German environmental regulations
Measuring principleZrO2 sensor with Pt-electrodes
Measuring range0 … 25 Vol%
Analogue signal port(0) 4 … 20 mA or 0 … 10V, galvanically separated
Reproducability+/- 2 % of the measuring value
ResolutionC(O2) – 0,01 %
T90-Time< 20 Seconds
Display2* 16 digit, illuminated LCD display
1. Line: display of concentration in Vol %
2. Line: messages, device status, sensor parameters
Messages1 System message (measuring value yes / no)
max. 2 messages configurable as measuring values
Digital communicationserial interface RS232
Ambient operation temp.+ 5 °C up to + 45 °C
Power supply24 / 110 / 230 VAC
Protection / Housing / DimensionsIP65 / wall mounting housing / ca. 300 x 320 x 130 mm (hxbxt)
IP20 / electronics unit 1/2 19" / 19" TU / 3HU
IP54 / portable housing 63 TU / 3 HU
ApprovalsTÜV certification 936/807023/A of 14.08.2000
Weight8 kg
OptionsAuto-calibration
back purge for high dust concentrations
pressure reducer
particle lfilter 2-7µm
manual 5 way valve
electronical/pneumatic gas supply pump
electronical flowmeter 
Software for remote control and service
Version: AMS 3220 E V-2013-07