The Oxygen Analyser AMS 3175 is the stationary measuring system for combustible and non combustible gases in Ex-classified areas.
The electronics and the gas analysis are separated from each other and installed in a gas-tight wall mounting housing. The AMS 3175 is Ex-certified for applications in Ex-Zone1, according Ex ia IIC T6.
Since the process gas is fed to the AMS Oxygen Analysers, usually by applying the pressure in the gas line on the analyser, it is essential to vent the gas at the outlet at atmospheric pressure. The diameter of the gas line at the outlet is 6 mm, and may not be reduced in diameter over the length of the vent-line.
Technical data
Analyser | AMS 3175 Ex |
---|---|
Ex-classification | ATEX 2 G, Ex ia IIC T6 |
Measuring principle | Electrochemical oxygen sensor |
Application | Gases Industries, Chemical Industries |
Measuring range | max. 4 manual selection 0 … 10, 0 … 100, 0 … 1000, 0 … 10000 ppmv |
Analogue signal port | 4 … 20mA two-wire transmitter |
Reproducability | +/– 2 % of the measuring value |
Resolution | 0,01 ppm – C(O2) – 1 ppm depending on the O2 concentration |
T90-Time | ca. 40 seconds |
Display | 4 ½ digit LCD display |
Messages | 2 free adjustable isolated changeover relays |
Gas connection | inlet / outlet 3 / 6 mm ferrule pack |
Gas sampling | built-in inlet / outlet valve flowmeter |
Sample flow | min. 20 Nl/h, max. 40 Nl/h |
Sample pressure (inlet) | min. 1,01 bar abs., max. 2 bar abs. |
Sample pressure (measuring cell) | max. 50 mbar pressure |
Ambient operating temp. | - 0 °C up to + 45 °C |
Relative humidity of the gas | 0 … 99 % not condensing |
Power supply | 24 VDC intrinsically safe |
Protection / Housing / | IP 65 / wall mounting housing off Aluminium die-casting, |
Dimensions | separated for electronic and gas analytics / 446 x 232 x 180 mm (hxbxt) |
Ex-classification | ATEX 2 G, Ex ia IIC T6 |
Weight | 9 kg |
Options | intrinsically safe power supply with signal separator Flame arrester ATEX II G IIC |
Version: AMS 3175 E V-2013-07 |